Answers are grounded in this paper and cite their sources.
OBJECTIVES: Zirconia ceramic was introduced for femoral heads because of its mechanical strength and fracture toughness. Highly cross-linked polyethylene has been shown to achieve wear reduction compared to conventional polyethylene. The purpose of this study was to investigate the long-term results of cementless total hip arthroplasty with zirconia femoral heads on highly cross-linked polyethylene. METHODS: Radiographic measurements of two-dimensional femoral head penetration into the polyethylene were performed with a computerized method (PolyWare) in 93 hips at a mean follow-up of 13 years. The steady-state wear rates were measured based on the radiographs from the first year after surgery to the final follow-up. Osteolysis was not found around the implant, and no zirconia heads had fractures. The total head penetration rate was 0.04 mm/year, and the steady-state wear rate was 0.02 mm/year. Kaplan-Meier survivorship with the endpoint of revision was 97% (95% confidence interval [CI], 92% to 99%) at 10 years. CONCLUSION: Zirconia on highly cross-linked polyethylene demonstrated extremely low wear properties at long-term follow-up.
No takes yet. Share an insight, caveat, or question.
Naito et al. (2020) studied this question.