Why the study?
Does the USCOM provide reliable cardiac output measurements compared to an aortic flowprobe in anesthetized dogs?
Does the USCOM provide reliable cardiac output measurements compared to an aortic flowprobe in anesthetized dogs?
The USCOM device provides reliable cardiac output measurements in a canine model, supporting the need for human clinical trials.
Answers are grounded in this paper and cite their sources.
Should not change human practice; leaves open translation of USCOM cardiac output monitoring via clinical trials.
We have used an animal model to test the reliability of a new portable continuous-wave Doppler ultrasonic cardiac output monitor, the USCOM. In six anesthetized dogs, cardiac output was measured with a high-precision transit time ultrasonic flowprobe placed on the ascending aorta. The dogs' cardiac output was increased with a dopamine infusion (0-15 microg x kg(-1) x min(-1)). Simultaneous flowprobe and USCOM cardiac output measurements were made. Up to 64 pairs of readings were collected from each dog. Data were compared by using the Bland and Altman plot method and Lin's concordance correlation coefficient. A total of 319 sets of paired readings were collected. The mean (+/-sd) cardiac output was 2.62 +/- 1.04 L/min, and readings ranged from 0.79 to 5.73 L/min. The mean bias between the 2 sets of readings was -0.0l L/min, with limits of agreement (95% confidence intervals) of -0.34 to 0.31 L/min. This represents a +/-13% error. In five of six dogs, there was a high degree of concordance, or agreement, between the 2 methods, with coefficients >0.9. The USCOM provided reliable measurements of cardiac output over a wide range of values. Clinical trials are needed to validate the device in humans.
No takes yet. Share an insight, caveat, or question.
Critchley et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: