Why the study?
Does the thermodilution technique provide comparable cardiac index measurements to the direct Fick method in patients following myocardial revascularization?
Does the thermodilution technique provide comparable cardiac index measurements to the direct Fick method in patients following myocardial revascularization?
Thermodilution provides cardiac index measurements that are reasonably close to the direct Fick method in most post-cardiac surgery patients, though discrepancies >0.5 L/min/m2 occur in about a quarter of cases.
Answers are grounded in this paper and cite their sources.
Thermodilution may support CI monitoring as a practical alternative post-revascularization; leaves open validation in broader or high-output cohorts.
With the thermodilution technique, repeated determination of cardiac index (CI) has become a rapid, easily accomplished, safe procedure. We paired 77 CI determinations measured by direct Fick and thermodilution techniques in 21 patients who underwent myocardial revascularization. Commercially available thermistor-tipped catheters and a cardiac output computer were used to determine CI according to the manufacturer's instructions. Oxygen consumption and arterial-venous oxygen content differences were measured directly to determine CI by the Fick method. Comparisons were made during mechanical ventilation with and without positive end-expiratory pressure (PEEP) and during spontaneous ventilation with and without PEEP. Cardiac indices measured by the two techniques were within +/- 0.5 L/min/m2 of each other only 76% of the time and within +/- 1.0 L/min/m2 96% of the time, if CI greater than or equal to 4.0 L/min/m2 were omitted. Ventilatory pattern had no apparent effect on results.
No takes yet. Share an insight, caveat, or question.
Hodges et al. (1975) studied this question.