Answers are grounded in this paper and cite their sources.
A nuclear magnetic resonance spectrometer has been constructed and its performance at 12 and 43 gauss is discussed. The magnetic field is obtained with an end corrected solenoid powered by batteries. The best resolution obtained is 10 microgauss at 12 gauss and 20 microgauss at 43 gauss over a 30 cm3 sample. A twin-T bridge and a synchronous detector are employed for the nuclear resonance detection. Some of the measurements for which this instrument is suited are described and the results of relaxation time measurements for ethanol-water mixtures are presented.