Does the Hatteland Doppler probe accurately detect fat emboli in an in vitro flow model?
Does the Hatteland Doppler probe accurately detect fat emboli in an in vitro flow model?
The tested Doppler method is insufficient for detecting fat emboli due to the low echogenicity of fat droplets.
Answers are grounded in this paper and cite their sources.
Doppler may miss most fat emboli in monitoring; leaves open whether refined probes or alternatives improve detection.
Pericardial suction blood (PSB) is known to be contaminated with fat droplets, which may cause embolic brain damage during cardiopulmonary bypass (CPB). This study aimed to investigate the possibility to detect fat emboli by a Doppler technique. An in vitro flow model was designed, with a main pump, a filter, a reservoir, and an injector. A Hatteland Doppler probe was attached to the circulation loop to monitor particle counts and their size distribution. Suspended soya oil or heat-extracted human wound fat was analyzed in the model. The concentrations of these fat emboli were calibrated to simulate clinical conditions with either a continuous return of PSB to the systemic circulation or when PSB was collected for rapid infusion at CPB weaning. For validation purpose, air and solid emboli were also analyzed. Digital image analysis was performed to characterize the nature of the tested emboli. With soya suspension, there was an apparent dose response between Doppler counts and the nominal fat concentration. This pattern was seen for computed Doppler output (p = .037) but not for Doppler raw counts (p = .434). No correlation was seen when human fat suspensions were tested. Conversely, the image analysis showed an obvious relationship between microscopy particle count and the nominal fat concentration (p < .001). However, the scatter plot between image analysis counting and Doppler recordings showed a random distribution (p = .873). It was evident that the Doppler heavily underestimated the true number of injected fat emboli. When the image analysis data were subdivided into diameter intervals, it was discovered that the few large-size droplets accounted for a majority of total fat volume compared with the numerous small-size particles (< 10 microm). Our findings strongly suggest that the echogenecity of fat droplets is insufficient for detection by means of the tested Doppler method.
No takes yet. Share an insight, caveat, or question.
Wikstrand et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: