The comparative performance of twelve different late-model color-measuring instruments was studied for precision (short-term repeatability) and conformance of color measurement and color-difference measurement.The instruments tested included four true and two abridged spectrophotometers and six colorimeters, all being integrating-sphere instruments except three 45 degrees /0 degrees colorimeters. All the colorimeters were individually calibrated for close conformance to CIE coordinates. Forty-four samples taken from the group studied in the previous paper were measured three times on each instrument, by three different well-trained operators. Again, a GE spectrophotometer was considered the reference instrument for conformance studies. Overall, the GE spectrophotometer, the Hunterlab D25D colorimeter, and the Kollmorgen KCS-18 and KCS-40 abridged spectrophotometers demonstrated the best performance. Statistical treatment of the data has been improved vastly since the earlier paper, and some recalculated results are included.