This case report highlights a rare and fortuitous event where a dislodged coronary stent embolized directly to a secondary target lesion and was successfully deployed.
Answers are grounded in this paper and cite their sources.
Stent dislodgement persists with modern balloon-mounted stents; leaves open optimal prevention and retrieval strategies in complex PCI.
Dislodgement of coronary stents while performing coronary interventions is not an uncommon phenomenon. Although more frequently encountered with stents that were manually crimped, dislodgement can also occur with the newer generation, balloon-mounted stents despite their superior trackability and deliverability. We describe a case with severe right coronary artery (RCA) and obtuse marginal (OM) disease in which an initial coronary angioplasty to the RCA followed by OM was planned. The RCA was stented and while negotiating a second balloon-mounted stent across the RCA, it slipped off the balloon and initially embolized into the aortic sinuses. Subsequently and fortuitously, the stent migrated into the OM vessel and was seated right at the lesion, which needed to be stented. Using the clear cell stent boost function, the embolized stent was wired and then deployed using sequential balloon dilations.
No takes yet. Share an insight, caveat, or question.
Senguttuvan et al. (2018) studied this question.