A new vectorcardiographic loop alignment method significantly reduces the confounding effects of respiration on morphologic beat-to-beat variability measurements.
Answers are grounded in this paper and cite their sources.
Loop alignment halves respiration-induced morphologic variability; leaves open clinical utility and optimal sampling rates pending validation.
The measurement of subtle morphologic beat-to-beat variations in the electrocardiogram is complicated by the presence of respiration-induced movements of the heart. A statistical signal model is developed which accounts for such movements by means of scaling, rotation, and time synchronization of vectorcardiographic loops. The maximum-likelihood estimator of the parameters describing these three transformations is presented and is extended to the case of multiple loop alignment. The performance of the method is assessed by measuring morphologic variability before and after loop alignment. It is shown that the effects of respiration on morphologic variability can be considerably reduced by the new method. Measurements on morphologic variability were typically reduced by a factor of 0.53 after loop alignment. The results show also that beat-to-beat measurements are strongly dependent on the selected sampling rate and that a rate of 1 kHz is too low.