Answers are grounded in this paper and cite their sources.
STUDY OBJECTIVES: The aim of this pilot study was to evaluate the clinical effectiveness of subjective titration versus objectively guided titration during polysomnography (PSG) and drug-induced sleep endoscopy (DISE) in mandibular advancement device (MAD) therapy for patients with obstructive sleep apnea (OSA). METHODS: In this pilot cross-over study, patients underwent three titration procedures in randomized order: (1) subjective titration, (2) PSG-guided titration using a remotely controlled mandibular positioner (RCMP) and (3) DISE-assisted titration using RCMP. After each titration procedure, patients used the MAD for 1 month at the targeted protrusion obtained according to the preceding titration procedure. For each procedure, a follow-up PSG was performed after 1 month of MAD use in order to evaluate the efficacy of the MAD. RESULTS: Ten patients were included in the study. Overall, no significant differences in targeted optimal protrusion compared to maximal comfortable protrusion among the three titration methods were observed. There was no significant difference in reduction in AHI. In this study, PSG titration correctly classified 50% of patients as 'responder'. A higher predictive accuracy was found for DISE titration with a sensitivity of 83.3% and a specificity of 100%. CONCLUSIONS: This pilot randomized cross-over trial showed no differences in optimal mandibular positioning and corresponding efficacy of MAD between subjective titration, DISE titration or PSG titration.
No takes yet. Share an insight, caveat, or question.
Kazemeini et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: