This is the authors' abstract. We don't add key points for this paper.
We demonstrate a microfabricated heat-flow calorimeter capable of measuring modulated heat currents with ∼5 pW resolution. This is achieved by combining the excellent thermal isolation of a microdevice suspended by thin and long beams (conductance ∼150 nW/K) with a high-resolution resistance thermometer that enables temperature measurements with 10–50 μK resolution [Sadat et al., Rev. Sci. Instrum. 83(8), 084902 (2012)). The calorimeter described here has a resolution comparable to state-of-the-art bimaterial cantilever-based calorimeters but surpasses previous designs by dissipating an order of magnitude lower power in the measurement process.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Sadat et al. (2013) studied this question.