Answers are grounded in this paper and cite their sources.
The polymerase chain reaction (PCR) is a standard molecular method that has the potential to solve the need for accurate, viable, and immediate infectious pathogen detection at point-of-care (POC) centers in different fields such as pathogen identification, and forensics. In this work, we present a photonic PCR thermocycler that achieves strong optical absorption, more efficient gene amplification, and further improves the effect of temperature distribution through the PCR sample using two gold nanofilms (AuNFs) and a high-power light-emitting diode chip (LED). The photonic device achieved higher heating and cooling rates of 13.20 null°C/s and 7.92 null°C/s, respectively on average, with uniform and reliable temperature distribution throughout the sample with negligible deviations. During amplification cycles, maximum temperatures attained had variations less than 1 °C at 90 °C, less than 0.8 °C at 55 °C, and less than 0.5 °C at 72 °C, showing uniform heating which yielded more accurate and reliable results. Using the fabricated device, PCR amplification for a bacteria genomic DNA was performed in 7.5 minutes for 30 thermal cycles with a sample volume of 20 μL. Due to the device's simple configuration and increased heat transfer rates, this photonic platform will be a highly efficient choice for rapid POC applications in developing countries.
No takes yet. Share an insight, caveat, or question.
Nabuti et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: