A novel ultrasound technique using 2D cross-correlation of RF data accurately measures perpendicular blood velocity components in a straight tube model.
Answers are grounded in this paper and cite their sources.
Supports precise experimental velocity mapping without averaging; leaves open clinical translation pending in vivo validation.
An ultrasound velocity assessment technique was validated, which allows the estimation of velocity components perpendicular to the ultrasound beam, using a commercially available ultrasound scanner equipped with a linear array probe. This enables the simultaneous measurement of axial blood velocity and vessel wall position, rendering a viable and accurate flow assessment. The validation was performed by comparing axial velocity profiles as measured in an experimental setup to analytical and computational fluid dynamics calculations. Physiologically relevant pulsating flows were considered, employing a blood analog fluid, which mimics both the acoustic and rheological properties of blood. In the core region (| r |/ a < 0.9), an accuracy of 3 cm s −1 was reached. For an accurate estimation of flow, no averaging in time was required, making a beat to beat analysis of pulsating flows possible.