Why the study?
Does proximal forearm non-invasive blood pressure measurement provide interchangeable mean arterial pressure readings compared to direct intra-arterial measurement in obese adults?
Does proximal forearm non-invasive blood pressure measurement provide interchangeable mean arterial pressure readings compared to direct intra-arterial measurement in obese adults?
Proximal forearm non-invasive blood pressure measurement is not interchangeable with direct intra-arterial measurement in obese patients due to wide dispersion of readings.
Answers are grounded in this paper and cite their sources.
Forearm NIBP cannot substitute for arterial MAP monitoring in obese patients; leaves open validation of alternative noninvasive sites.
Background: In obesity, accurate perioperative blood pressure measurement using upper arm, non-invasive blood pressure (NIBP) is technically challenging. Proximal forearm NIBP may be an acceptable substitute. Mean arterial blood pressures (MAP) estimated by proximal forearm NIBP were compared with direct intra-arterial measurements. It was hypothesised that the measurement techniques would be interchangeable if between-technique MAP differed ≤ 20% and MAP ratios were < 1.2 and > 0.8.Method: A total of 30 adults with body mass index ≥ 30 kg/m2 in whom perioperative intra-arterial blood pressure measurement was considered mandatory were enrolled. MAP measurements using the two techniques were obtained at three random intervals in each patient. Bland–Altman analyses were employed.Results: Forearm mean NIBP MAP overestimated mean intra-arterial MAP by 2.2 (SD 8.1; range from 23.8 to –19.4 mmHg; p = 0.011, 95% CI 3.9 to 0.5). However, Bland–Altman analyses revealed a wide dispersion with several MAP differences and MAP ratios exceeding the pre-specified bounds for interchangeability.Conclusion: Forearm NIBP could not be considered interchangeable with direct intra-arterial MAP measurements in obese patients.
No takes yet. Share an insight, caveat, or question.
Verkhovsky et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: