Does a new wireless PC-assisted analog device accurately measure vectorcardiography compared to a standard electrocardiograph in volunteers?
Does a new wireless PC-assisted analog device accurately measure vectorcardiography compared to a standard electrocardiograph in volunteers?
A newly developed low-cost wireless PC-assisted device for vectorcardiography demonstrates acceptable accuracy with maximum uncertainties of 20% in cardiac axis orientation.
Answers are grounded in this paper and cite their sources.
Acceptable for preliminary vectorcardiography research; leaves open clinical use pending larger validation studies.
The development of an analog device to wirelessly modulate and record twelve cardiac leads, PC assisted is reported. The system, which has been developed using low cost components, contains only one signal conditioning stage and one stage for amplitude modulation with FM transmission/reception. Also it contains an interface to select the cardiac leads remotely, which is based in the transmission/decoding of audible dual frequency tones. The device has been tested on a group of volunteers simultaneously using a registered trademark electrocardiograph, to assess the performance of the new device. Furthermore, the experimental results indicate that the device is a tool suitable to conduct studies of vector-cardiography with maximum uncertainties of 20 % in the cardiac axis orientation measurement.