This is the authors' abstract. We don't add key points for this paper.
This work proposed a novel 4-channel wireless data logger that uses electromyogra- phy (EMG), electrocardiograph (EKG), and accelerometer transducers for field measurements of biosignals. The logger, which weighs 102 g (including the battery), continually acquires 4-channel 16-bit analog signals at 1000 Hz/channel. The acquired data are saved on a microSD memory card for subsequent analysis. The logger utilizes a ZigBee transceiver module that transmits acquired data to a remote receiver. The logger can run for over 6 consecutive hours at full load when powered by two 860 mAh batteries. The remote receiver unit, connected to a PC or laptop via a USB interface, was controlled by Window-based monitoring software for com- municating with the logger. The monitoring software programmed by Borland C++ Builder op- erated the logger and monitored real-time data sent from the logger. Three laboratory experi- ments and two prolonged field tests were performed to test logger feasibility. The experimental results demonstrate that the system is reliable and feasible for field measurements of biosignals for a prolonged period.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2010) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: