Answers are grounded in this paper and cite their sources.
This paper describes the use of germanium lithium-drift p-i-n diodes as high-resolution γ-ray spectrometers. With these spectrometers we have obtained γ-ray resolutions of 2.05 keV at 122 keV, 4.0 keV at 1333 keV, and 5.5 keV at 2614 keV. Using the detectors as pair spectrometers for high-energy γ rays, we have obtained a resolution of 9.8 keV on a 7.6-MeV γ ray. The factors affecting the resolution of the detectors are discussed. Fano factors of ~0.4 have been observed. Efficiency curves are given for a 2.5 cm 2 × 3.5 mm detector and for a 5 cm 2 × 8 mm detector.The detectors have been used to make high-resolution studies of the complex γ-ray spectra from sources of 131 Cs, 161 Pm, 153 Gd, 156 Eu, 159 Gd, 177 Yb, and 226 Ra. Results are reported for the energies and intensities of the γ rays observed in these studies.
No takes yet. Share an insight, caveat, or question.
Ewan et al. (1964) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: