Key result
Percutaneous closure of a missed atrial septal defect using an Amplatzer device in a 67-year-old man with recurrent inappropriate ICD shocks led to an uneventful 3-year follow-up.
Why the study?
Inappropriate shocks occur in approximately one-third of patients with ICDs, and this report describes an unusual case caused by AF secondary to a missed ASD following a modified Bentall procedure.
Case Report (n=1)
Comprehensive pre- and post-operative echocardiographic assessment is crucial to identify missed structural abnormalities like an ASD, which can cause atrial fibrillation and inappropriate ICD shocks.
May support ASD closure to avert inappropriate ICD shocks in select cases; hypothesis-generating and requires prospective validation.
BACKGROUND: Inappropriate shocks have been reported in ∼1/3 of patients with implantable cardiac defibrillators (ICDs). We report an unusual case of inappropriate ICD shocks due to atrial fibrillation (AF) caused by a missed atrial septal defect (ASD) in a patient with a modified Bentall procedure. CASE SUMMARY: A 67-year-old Caucasian male, with an ICD and a history of a modified Bentall procedure 24 years ago, reported to our outpatient clinic with recurrent inappropriate ICD shocks due to episodes of fast AF. The transthoracic echocardiographic exam revealed two large aneurysms at the ostia of the coronary arteries. We performed further evaluation with transoesophageal echocardiogram and computed tomography (CT) angiography. The aneurysms measured on CT were 3.14 cm × 2.29 cm on the right ostium and 1.9 cm × 0.99 cm on the left. A large secundum type ASD of 1.5 cm, missed in all previous echocardiographic studies, was revealed. The therapeutic options of surgical closure of the ASD and repair of the aneurysms or a more conservative approach with percutaneous closure of the ASD and closer follow-up were discussed with the patient. The patient declined the surgical option due to high complication risk, and closure of the ASD with an Amplatzer device was performed 3 months later. A 3-year follow-up was uneventful. CONCLUSION: It is of major importance to comprehensively and thoroughly assess patients before and after a surgical intervention to not miss other treatable conditions preoperatively and complications in the postoperative period.
No takes yet. Share an insight, caveat, or question.
Κωστοπούλου et al. (2021) conducted a case report in Inappropriate ICD shocks, atrial fibrillation, missed atrial septal defect (n=1). Percutaneous closure of ASD with an Amplatzer device was evaluated on Clinical outcome. Percutaneous closure of a missed atrial septal defect using an Amplatzer device in a 67-year-old man with recurrent inappropriate ICD shocks led to an uneventful 3-year follow-up.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: