Key result
High-power short duration contact force ablation catheters significantly shortened total lesion time by 758.3 seconds (95% CI -1128.29 to -388.35) compared to 8 mm tip catheters.
Why the study?
Although typical atrial flutter can be easily ablated with radiofrequency energy, studies comparing the effectiveness of different ablation catheters were lacking.
Does the type of radiofrequency ablation catheter affect ablation success and lesion time in patients with typical atrial flutter?
Observational (n=222)
No
Does the type of radiofrequency ablation catheter affect ablation success and lesion time in patients with typical atrial flutter?
Mean Difference: -758.3 (95% CI -1128.29–-388.35)
Contact force ablation catheters using high-power short-duration settings significantly reduce lesion time during typical atrial flutter ablation without compromising the high procedural success rate.
May shorten ablation times; leaves open impact on efficacy, safety, and outcomes in randomized trials.
Introduction Although ablation of typical atrial flutter (AFL) can be easily achieved with radiofrequency energy (RF), there are no studies that compare effectiveness of different ablation catheters. Our study aimed to compare the effectiveness of various types of ablation catheters in the treatment of AFL. Methods We analysed patients with AFL who underwent RF ablation by a single operator at our institution. Successful ablation was evidenced by presence of bidirectional conduction block (trans-isthmus conduction time ≥130 ms, or doubling of baseline conduction time, or presence of double potentials ≥90ms). Logistic regression was used to compare success rate and linear regression to compare lesion time. Results Out of the 222 patients, only 6 patients did not meet success criteria (2.7%). Catheters used were 8 mm tip in 16 patients, internally irrigated (Chili II Boston Scientific) in 47 patients, externally irrigated (non-force sensing) catheters (CoolPath, Abbott) in 40 patients. Externally irrigated force sensing catheter (Tacticath, Abbott) was used with >10 gm of force and (LPLD) setting (30W-45°C-60 sec) in 50 patients, and high-power short duration (HPSD) setting (50W-43°C −12 sec,) in 70 patients. No complications were encountered. Catheter type had no statistically significant association with ablation success. In terms of lesion time, HPSD catheter statistically significantly shortened lesion time by 758.3s, [CI −1128.29, −388.35s] followed by LPLD by 419.0s [CI −808.49, −29.47s]. Table 1 shows the lesion time difference for the catheters used as compared with 8 mm tip. Conclusions Typical atrial flutter radiofrequency ablation procedure had a high success rate, not influenced by type of ablation catheter. Contact force ablation catheter on HPSD is associated with shorter total lesion time. Funding Acknowledgement Type of funding sources: None.
No takes yet. Share an insight, caveat, or question.
Gorantla et al. (2022) conducted an observational in Typical atrial flutter (n=222). High-power short duration (HPSD) contact force ablation catheter vs. 8 mm tip catheter was evaluated on Lesion time (MD -758.3s, 95% CI -1128.29 to -388.35). High-power short duration contact force ablation catheters significantly shortened total lesion time by 758.3 seconds (95% CI -1128.29 to -388.35) compared to 8 mm tip catheters.