Authors
The artificial pancreas (AP) has been tested extensively in the hospital setting (1–5). Here we describe a next logical step in AP development—the first outpatient trials of a wearable AP based on a smartphone computational platform. Following Ethical Committee approvals and ClinicalTrials.gov registration (NCT01447992 and NCT01447979), two simultaneous studies were conducted in Padova, Italy, and Montpellier, France, in October 2011, enrolling a 38-year-old female and a 52-year-old male, respectively; both were Caucasian, type 1 diabetic insulin pump users. Day 1—At 17:00, the patients arrived at hotels located within 1 km from the emergency room. Subjects’ pumps were replaced by Omnipod Insulin Management Systems. The APs were activated in open-loop mode implementing the patients’ regular routines and remote monitoring was initiated. At 20:00, the patients had dinner at a local restaurant, without dietary restrictions and then spent the night in the hotel. Day 2—At 7:00, the patients were admitted to the clinic and the APs were switched to automated closed-loop control and challenged by breakfast at 8:00 and lunch at 12:00. At 18:00, the patients moved back to the hotel; dinner was at 20:00 in a local restaurant, without dietary restrictions. Meal bolus was recommended by the APs and approved by the patients; …
Loading...
Cobelli et al. (2012) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: