Key result
Trained observer auscultatory measurements correlated significantly better with daytime ambulatory diastolic blood pressure than the automated Omron 705CP device (r=0.593 vs 0.319, P<0.0001).
Why the study?
Does blood pressure measurement by trained observers correlate better with daytime ABPM than an automated oscillometric device in untreated hypertensive patients?
Observational (n=313)
Does blood pressure measurement by trained observers correlate better with daytime ABPM than an automated oscillometric device in untreated hypertensive patients?
Absolute Event Rate: 0.593% vs 0.319%
p-value: p=<0.0001
Trained observers using auscultatory methods provide blood pressure measurements that correlate better with daytime ABPM than the Omron 705CP automated device, particularly for diastolic blood pressure.
Auscultatory measurements may warrant preference in BP research; leaves open whether automated devices suffice for clinical decision-making.
OBJECTIVES: Accurate blood pressure measurement is critical to successful clinical trials. Concerns about observer errors have led to the use of automated oscillometric devices without evidence that their performance is similar to that of trained observers. This study compares blood pressures obtained by trained observers and with an oscillometric device (Omron 705CP) to 24-h ambulatory blood pressure monitoring. METHODS: We performed a post-hoc analysis of 313 untreated hypertensive patients at the end of the washout phase of a Novartis hypertension trial. Patients had three seated trained observer mercury auscultatory blood pressure measurements followed by 24-h ambulatory blood pressure monitoring. The next day, the ambulatory blood pressure monitoring was removed and three seated readings were obtained with an Omron 705CP. Correlations for systolic blood pressure and diastolic blood pressure were obtained between daytime ambulatory blood pressure monitoring (0900 and 2100) and the two office methods. In addition, we investigated the degree of difference of trained observer and Omron measurements from ambulatory blood pressure monitoring. RESULTS: For systolic blood pressure, the correlation with ambulatory blood pressure monitoring of the trained observer was significantly better than with that of the Omron 705CP (0.641 vs. 0.555, P=0.01). For diastolic blood pressure values, even greater disparity between the two office method correlations with ambulatory blood pressure monitoring was observed (trained observer=0.593 vs. Omron=0.319, P<0.0001). Both trained observer and Omron readings were consistently higher than ambulatory blood pressure monitoring for systolic blood pressure (P<0.0001) and diastolic blood pressure (P<0.0001). Omron measurements, however, deviated from ambulatory blood pressure monitoring more than those of the trained observer (P<0.0001 for systolic blood pressure and diastolic blood pressure). CONCLUSIONS: For clinical trials using diastolic blood pressure targets, the Omron 705CP cannot replace the auscultatory blood pressure measurements of a trained observer. For systolic blood pressure, the Omron device and the trained observer had similar correlations with ambulatory blood pressure monitoring; however, both methods gave consistently higher systolic blood pressure values. Further study of oscillometric devices should be conducted before universally replacing auscultatory blood pressure determinations by trained observers in clinical trials.
No takes yet. Share an insight, caveat, or question.
Graves et al. (2006) conducted an observational in Hypertension (n=313). Trained observer auscultatory blood pressure measurement vs. Automated oscillometric device (Omron 705CP) was evaluated on Correlation of diastolic blood pressure with daytime ambulatory blood pressure monitoring (p=<0.0001). Trained observer auscultatory measurements correlated significantly better with daytime ambulatory diastolic blood pressure than the automated Omron 705CP device (r=0.593 vs 0.319, P<0.0001).