One hundred ninety-three geometric total knee arthroplasties (TKA) were performed between 1972 and 1975 in 129 patients (66 women, 63 men; mean age, 69 years) with osteoarthritis. Of these, 102 knees were followed for a mean of 11 years. Eighty-three percent of the patients had mild or no pain. The revision rate was 20%, and the surgical complication rate was 12%. By actuarial analysis, the probability of retaining a geometric prosthesis at 10 years was 78%. With revision or moderate to severe pain as the end point, the predicted implant survival was 69% at 10 years. Lucent lines greater than 1 mm were present in 38% of the knees and progressed in 34%; they were more frequent in knees with greater than or equal to 3 degrees of varus axial alignment (p less than 0.05) or greater than or equal to 4 degrees of varus placement of the tibial component (p less than 0.05). The geometric prosthesis has provided a functional result in 69% of knees at 10 years, despite being the first two-part component knee replacement retaining the cruciate ligaments and using early surgical instrumentation and implant design.
No takes yet. Share an insight, caveat, or question.
Rand et al. (1988) studied this question.