Answers are grounded in this paper and cite their sources.
Several modifications of GC-MS and HPLC methods for plasma level DHPG have been described. The effects of storage temperature and stabilizing agents on DHPG stability have been studied. The stabilizing agent has been found to play a more important role than low-temperature storage in preventing DHPG from decomposition during sample storage. A specific and sensitive GC-MS method (electron impact) has been established using stable isotope-labeled DHPG as an internal standard. HPLC has been improved by modifying the conditions, resulting in a good separation of DHPG and internal standard from solvent front other early eluting compounds. Comparison of the GC-MS and HPLC procedures demonstrates a strong correlation between these two methods.
No takes yet. Share an insight, caveat, or question.
Xie et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: