Key result
A monitoring bracelet recorded longer walking distance distances compared to the standard method in stable heart failure patients.
Why the study?
Does a monitoring bracelet accurately measure 6-minute walk distance and heart rate compared to standard methods in stable heart failure patients?
Observational (n=88)
No
Does a monitoring bracelet accurately measure 6-minute walk distance and heart rate compared to standard methods in stable heart failure patients?
Answers are grounded in this paper and cite their sources.
Monitoring bracelets may overestimate 6-minute walk distance compared to standard methods in heart failure patients, although heart rate measurements are comparable.
Objective To investigate the feasibility of monitoring bracelet in 6-minute walk test(6MWT). Methods Stable heart failure patients hospitalized in Heart Failure Care Unit of Fuwai Hospital from July 2016 to December 2016 were enrolled in our study. Dynamic telemetry ECG recording monitor and monitoring bracelet were used in 6MWT meanwhile , recording resting heart rate(RHR) and immediate heart rate after 6MWT, respectively. Monitoring bracelet also measured walk distance to compare with the distance in the standard method. Result 88 patients were included in our study, with age of (50.22±8.85) years and a male proportion of 81%(71/88). The 6-minute walk distance recorded by monitoring bracelet was significantly longer than standard method[(442.85±81.00)m vs (377.91±86.13)m, P<0.01]. There was no significant difference in heart rate before and after 6MWT[RHR: (73.14±10.26)bpm vs (74.28±10.40)bpm, P=0.587; Immediate HR after 6MWT: (87.53±13.68)bpm vs (89.09±13.82)bpm, P=0.162]. Conclusion 6-minute walk distance measured by monitoring bracelet is higher than standard measurement, which may overestimate heart function of heart failure patients. In terms of heart rate measurement, monitoring bracelet can provide reference for heart failure patients to perform home-based cardiac rehabilitation in the future.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2018) conducted an observational in Stable heart failure (n=88). Monitoring bracelet vs. Standard method was evaluated on 6-minute walk distance (p=<0.01). A monitoring bracelet recorded longer walking distance distances compared to the standard method in stable heart failure patients.
Synapse has enriched 10 related papers. Consider them for comparative context: