Answers are grounded in this paper and cite their sources.
Dynamic adaptation of the linearity criterion improves powerline interference subtraction in ECG and other biosignals, especially in the presence of EMG noise.
The performance of a previously developed and widely used method for powerline interference subtraction from the ECG is slightly reduced in the presence of continuous well-expressed EMG noise. Applying automatic adaptation of the 'linearity criterion' value in accordance to the ECG/noise ratio, the best conditions for application of the procedure can be obtained. The proposed method allows for reduction of possible distortions when applied on noise-free ECG and enhances its efficiency in the presence of non-powerline noise. Apart from ECG, it is applicable to impedance-cardiogram, plethysmogram, EEG and other biosignals.
No takes yet. Share an insight, caveat, or question.
Ivaylo Christov (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: