This is the authors' abstract. We don't add key points for this paper.
This study aimed to evaluate cyclic fatigue resistance and bending properties and torque/force generation of WaveOne Gold (WOG) Primary in comparison with WaveOne (WO) Primary and Reciproc (RE) R25. A cyclic fatigue test revealed that the WOG Primary took significantly longer time to fracture compared with the WO Primary (p<0.05). The WOG Primary had the smallest load values at a deflection of 0.5 and 2 mm (p<0.05), as measured with a cantilever bending test. Torque/force measurement demonstrated that maximum upward force and maximum counterclockwise torque values in the WOG Primary were significantly lower than those in the RE R25 (p<0.05). Under the present experimental condition, the WOG Primary showed a higher cyclic fatigue resistance compared with the WO Primary, a higher flexibility compared with the WO Primary and RE R25, and generated a significantly lower maximum torque compared with the RE R25.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Tong et al. (2019) studied this question.