A novel carbon fiber-based method enables the precise measurement of force and shortening in single ventricular myocytes under various load conditions, facilitating cellular-level cardiac mechanics research.
Answers are grounded in this paper and cite their sources.
Suggests optimal myocyte compliance for work output; hypothesis-generating for cellular mechanics but leaves in vivo relevance open.
To facilitate cardiac muscle research, we developed a novel method by which the force and length of a single ventricular myocyte can be recorded with a pair of carbon graphite fibers attached firmly to both ends. One fiber was stiff, whereas the other fiber was compliant to allow the recording of force and shortening during twitch contractions. The image of the compliant carbon fiber was projected onto a pair of photodiodes, and their output was fed to a piezoelectric transducer after variable amplifications to alter the effective compliance of the carbon fiber. Thus contraction of the myocyte was induced under virtually isometric conditions as well as under auxotonic conditions. We obtained a bell-shaped relation between the compliance under an auxotonic load and the work output of the myocyte, which was directly related to myocyte performance in the heart. Because it is easy to attach myocytes to the experimental apparatus, the present method would allow us to study cardiac muscle mechanics at the cellular and molecular levels.
No takes yet. Share an insight, caveat, or question.
Yasuda et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: