Answers are grounded in this paper and cite their sources.
The design, fabrication, and performance of miniature high-frequency annular arrays are described. A 50-MHz, 2-mm-diameter, 7-element, equal-area annular array was fabricated and tested. The array elements were defined using photolithography and the electrical contacts were made using ultrasonic wire bonding. The resulting transducer produced pulses with a -6 dB bandwidth of 52% and an insertion loss of -16 dB. A radiation pattern was collected by scanning the transducer array above the tip of a glass fiber. A -6 dB two-way beam width of 75 microns was found at f/2. The radiation pattern decreased smoothly to less than -60 dB at a distance of 550 microns.
No takes yet. Share an insight, caveat, or question.
Brown et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: