Why the study?
Pacing threshold increases shortly after Micra transcatheter pacing system implantation in some patients, leading to battery depletion; this study aimed to define the causes and predictors of this increase.
What are the predictors of an increased pacing threshold shortly after Micra transcatheter pacing system implantation?
What are the predictors of an increased pacing threshold shortly after Micra transcatheter pacing system implantation?
Implant impedance >660 Ω and threshold <1.0 V/0.24 msec may predict a stable pacing threshold and prevent micro-dislodgement after Micra-TPS implantation.
Early threshold rise after Micra may warrant closer monitoring; this Level 5 analysis leaves open optimal preventive strategies.
Introduction: Achievement of a favorable pacing threshold with a Micra transcatheter pacing system (Micra-TPS) is important for reducing battery depletion; in some cases, the threshold increases shortly after the device is implanted, and a higher pacing threshold may be required. This study aimed to define the causes and predictors of the increase in pacing threshold observed shortly after Micra-TPS implantation. Methods and Results: The study included 64 consecutive patients who underwent Micra-TPS implantation between 2017 and 2020. Patients were divided into two groups depending on their pacing threshold, namely, the increased pacing threshold (IPT) group (threshold increased by ≥0.5 V/0.24 msec within 1 month of implantation) and the stable pacing threshold (SPT) group. Excluding 4 patients we were unable to conduct follow-up on, 9 of the 60 remaining patients (15%) were in the IPT group and 51 (85%) were in the SPT group. The IPT group had significantly lower implant impedance values and higher implant thresholds than the SPT group: 582 ± 59 vs 755 ± 167 Ω (P <0.001) and 1.29 ± 0.87 vs 0.71 ± 0.40 V/0.24 msec (P =0.014), respectively. Implant impedance and threshold may serve as predictors of a threshold increase after implantation (area under the curve: 0.737–0.943 and 0.586–0.926, respectively). Conclusion: An increased pacing threshold was noted shortly after Micra-TPS implantation owing to micro-dislodgement because of insufficient anchoring of the device on the myocardium. Impedance >660 Ω and threshold <1.0 V/0.24 msec may be predictive factors for an increased pacing threshold.
No takes yet. Share an insight, caveat, or question.
Higuchi et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: