Key result
Retrieval of a chronically implanted Micra transcatheter pacing system after 29 months was successfully performed using a snare loop, followed by implantation of a new leadless pacemaker.
Why the study?
The feasibility and safety of retrieving a chronically implanted Micra transcatheter pacing system had not been demonstrated in humans.
Case Report (n=1)
Demonstrates the feasibility of retrieving a chronically implanted leadless pacemaker (29 months) using a snare loop.
Answers are grounded in this paper and cite their sources.
Retrieval feasible after 29 months in this case; leaves open generalizability, safety, and need for prospective data before wider adoption.
We report the first-in-human retrieval of a chronically implanted Micra transcatheter pacing system (TPS; Medtronic, Minneapolis, MN, USA). A 41-year-old woman suffering from third-degree atrioventricular block was successfully implanted with a TPS in the low septum of the right ventricle. After 29 months, due to an increased electrical threshold and 100% pacing percentage, the device showed low battery voltage. Despite the long implant duration, the attempt of device retrieval using a snare loop inserted in the delivery system of a TPS was successful. In the same procedure, a new leadless pacemaker was implanted in the high right ventricular septum with optimal pacing threshold.
No takes yet. Share an insight, caveat, or question.
Curnis et al. (2019) conducted a case report in Third-degree atrioventricular block (n=1). Retrieval of chronically implanted Micra transcatheter pacing system was evaluated on Successful device retrieval and implantation of a new leadless pacemaker. Retrieval of a chronically implanted Micra transcatheter pacing system after 29 months was successfully performed using a snare loop, followed by implantation of a new leadless pacemaker.