Answers are grounded in this paper and cite their sources.
Experimental study reveals linear growth of electric current with acoustic decibel increases in a cochlear implant user, supporting logarithmic signal processing transformations.
Estimates of loudness balance were obtained for acoustically and electrically presented 250 Hz sine signals from a patient who uses the Ineraid multichannel cochlear implant. Acoustic and electric loudness matching was possible because the patient evidenced a 25 dB HL threshold at 250 Hz in his nonimplanted ear. The level of the electrical stimulus in microamperes required for a balance of loudness grew linearly with equal increments in decibels for the acoustic stimulus. These data, in concert with the very limited data from previous studies, provide a rationale for using a logarithmic transformation of acoustic to electric intensity in signal processors for cochlear implants.
No takes yet. Share an insight, caveat, or question.
Dorman et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: