Key result
Strategic ICM reprogramming linked to ~98% fewer false-positive alerts.
Observational (n=78)
Answers are grounded in this paper and cite their sources.
AIMS: Implantable cardiac monitors (ICMs) can generate false-positive (FP) alerts. Although these devices have an extended programmability, there are no recommendations on their optimization to reduce not-relevant activations.We tested a strategic programming optimization guide based on the type of FP and investigated the safety and feasibility of the nurse-led insertion of ICMs with a long-sensing vector. METHODS AND RESULTS: Consecutive patients implanted by trained nurses with long-sensing vector ICM were enrolled in a 1-month observational stage (Phase A). Patients who had ≥10 FP episodes underwent ICM reprogramming based on the predefined guide and were followed for an additional month (Phase B). A total of 78 patients had successful ICM insertion by nurses with a mean R wave amplitude of 0.96 ± 0.43 mV and an 86% P wave visibility. Only one patient reported a significant device-related issue, and nurse-delivered ICM was generally well accepted by the patients. During Phase A, 11 patients (14%) generated most of FP (3,627/3,849; 94%) and underwent ICM reprogramming. In the following month (Phase B), five patients (45%) were free from FP and six (55%) transmitted 57 FP alerts (98% reduction compared with Phase A). The median number of FP per patient was significantly reduced after reprogramming [195 (interquartile range, 50-311) vs. one (0-10), P = 0.0002]. CONCLUSION: A strategic reprogramming of ICM in those patients with a high FP alert burden reduces the volume of erroneous activations with potential benefits for the remote monitoring service. No concerns were raised regarding nurse-led insertion of ICMs with a long-sensing vector.
No takes yet. Share an insight, caveat, or question.
Maines et al. (2022) conducted an observational in Patients with implantable cardiac monitors (ICMs) and false-positive remote alerts (n=78). Strategic ICM reprogramming optimization guide and nurse-led insertion of long-sensing vector ICM vs. Pre-reprogramming baseline (Phase A) was evaluated on Median number of false-positive (FP) alerts per patient (p=0.0002). Strategic reprogramming of implantable cardiac monitors reduced median false-positive alerts per patient from 195 to 1 (98% overall reduction; P=0.0002).
Synapse has enriched 10 related papers. Consider them for comparative context: