Answers are grounded in this paper and cite their sources.
To compare the effectiveness of chest dual-energy subtraction digital tomosynthesis (DES-DT) with that of conventional digital tomosynthesis for the detection of calcifications superimposed over simulated pulmonary nodules. A DES-DT system with pulsed X-rays and rapid kV switching was used to examine calcifications in simulated pulmonary nodules. Low-voltage, high-voltage, and soft-tissue or bone-subtracted tomograms of the desired layer thicknesses were reconstructed from the image data acquired during a single tomographic scan, bone-subtracted images, and a scan angle of 40. Our analysis took into account the signal-to-noise ratio (SNR) of the different degrees of calcification in the simulated pulmonary nodules. For DES-DT, the SNR for the simulated pulmonary nodules increased about 63%. Based on the results of receiver operating characteristic performance analysis, the detection ability of our DES-DT was significantly better than that of conventional digital tomosynthesis (P < 0.03). The study results confirmed with a further study to assess influence of reconstruction and filtering for detection of simulated nodules. DES-DT provided greater sensitivity than conventional digital tomosynthesis.
No takes yet. Share an insight, caveat, or question.
Tsutomu Gomi (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: