The variable line spacing plane grating monochromator beamline at the Canadian Light Source (CLS) employs three grazing incidence variable line spacing gratings to cover a photon energy range of 5-250 eV. It uses a 185 mm period length planar permanent magnet insertion device as the photon source, sharing a straight section with another soft x-ray beamline at the CLS. The commissioning and performance of the beamline is reported. The high resolution photoabsorption spectra of Ar and PF(5) gases are reported. A resolving power of over 40,000 for photons in the low energy region and >10,000 for a wider energy range (8-200 eV) can be achieved. A photon flux of up to 2 x 10(12) photons/s per 100 mA with slit settings of 50 microm has been measured.