Why the study?
Right-sided ICD can placement may increase defibrillation threshold due to suboptimal shock vectors, and whether alternate RV coil positioning or adding SVC and CS coils can mitigate this was unknown.
Does the addition of SVC and CS coils or alternate RV coil positioning mitigate elevated defibrillation thresholds in right-sided ICD generator placement in a simulation model?
Does the addition of SVC and CS coils or alternate RV coil positioning mitigate elevated defibrillation thresholds in right-sided ICD generator placement in a simulation model?
In a simulation model, the elevated defibrillation threshold associated with right-sided ICD placement can be mitigated by using an apical RV shock coil and adding SVC and CS coils.
Right-sided ICD cans may raise DFT; modeling supports additional coils to mitigate but leaves open clinical adoption.
AIMS: The standard implantable cardioverter defibrillator (ICD) generator (can) is placed in the left pectoral area; however, in certain circumstances, right-sided cans may be required which may increase defibrillation threshold (DFT) due to suboptimal shock vectors. We aim to quantitatively assess whether the potential increase in DFT of right-sided can configurations may be mitigated by alternate positioning of the right ventricular (RV) shocking coil or adding coils in the superior vena cava (SVC) and coronary sinus (CS). METHODS AND RESULTS: A cohort of CT-derived torso models was used to assess DFT of ICD configurations with right-sided cans and alternate positioning of RV shock coils. Efficacy changes with additional coils in the SVC and CS were evaluated. A right-sided can with an apical RV shock coil significantly increased DFT compared to a left-sided can [19.5 (16.4, 27.1) J vs. 13.3 (11.7, 19.9) J, P < 0.001]. Septal positioning of the RV coil led to a further DFT increase when using a right-sided can [26.7 (18.1, 36.1) J vs. 19.5 (16.4, 27.1) J, P < 0.001], but not a left-sided can [12.1 (8.1, 17.6) J vs. 13.3 (11.7, 19.9) J, P = 0.099). Defibrillation threshold of a right-sided can with apical or septal coil was reduced the most by adding both SVC and CS coils [19.5 (16.4, 27.1) J vs. 6.6 (3.9, 9.9) J, P < 0.001, and 26.7 (18.1, 36.1) J vs. 12.1 (5.7, 13.5) J, P < 0.001]. CONCLUSION: Right-sided, compared to left-sided, can positioning results in a 50% increase in DFT. For right-sided cans, apical shock coil positioning produces a lower DFT than septal positions. Elevated right-sided can DFTs may be mitigated by utilizing additional coils in SVC and CS.
No takes yet. Share an insight, caveat, or question.
Qian et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: