Answers are grounded in this paper and cite their sources.
An improved calibration technique for an airborne Lyman-alpha hygrometer is presented. Like previous methods, it relies upon simultaneous measurement of absolute humidity determined from a slower response hygrometer. We show that a substantial improvement in the Lyman-alpha calibration is obtained by accounting for the time lag of the slower instrument. To show our technique we use data from Lyman-alpha and thermoelectric devices on the NCAR Electra during an investigation of the nearly neutral boundary layer over the Arabian Sea as part of the WMO/ICSU Summer Monsoon Experiment. We also show that for near-neutral conditions the eddy-correlation water vapor flux can be adequately estimated using the fast response vertical velocity data from a gust probe and slower response data from the thermoelectric device, which has been properly advanced to account for the time lag.
No takes yet. Share an insight, caveat, or question.
Friehe et al. (1986) studied this question.