BACKGROUND: Handheld ultrasound (HHUS) devices have chiefly been deployed in emergency medicine, where they are considered a valid tool. The data situation is less clear in the case of internal questions in abdominal sonography. In our study, we investigate whether HHUS devices from different manufacturers differ in their B-scan quality, and whether any differences are relevant for the significance of an internal ultrasound examination. METHOD: The study incorporated eight HHUS devices from different manufacturers. Ultrasound videos of seven defined sonographic questions were recorded with all of the devices. The analogue recording of the same findings with a conventional high-end ultrasound (HEUS) device served as an evaluation criterion. Then, the corresponding findings were played side by side and evaluated by fourteen ultrasound experts using a point scale (5 points = very good; 1 point = insufficient). Regardless of the manufacturer, the HHUS devices performed best when determining the status/inferior vena cava volume and in the representation of ascites/free fluid. CONCLUSION: In various clinical abdominal sonography questions, HHUS systems can reliably reproduce findings, and are-while bearing their limitations in mind-an acceptable alternative to conventional HEUS systems. Irrespective of this, the present study demonstrated relevant differences in the B-scan quality of HHUS devices from different manufacturers.