Key result
A rapid liquid chromatography and time-of-flight mass spectrometry assay for flecainide was validated, demonstrating linearity over 100-/L with a total analysis time of 1.2 minutes.
Why the study?
Flecainide measurement is needed to optimize dosing and avoid toxicity, particularly for urgent sample analysis during transplacental therapy for fetal tachycardia.
A rapid, 1.2-minute LC-TOF-MS assay was successfully validated for therapeutic drug monitoring of flecainide in human plasma.
Answers are grounded in this paper and cite their sources.
May enable faster flecainide TDM in labs; leaves open clinical outcome benefits versus existing assays.
BACKGROUND: Measurement of flecainide is useful to optimize dosage and minimize risks of toxicity. Furthermore, there is a need for urgent sample analysis when flecainide is used in transplacental therapy for fetal tachycardia. To this end, we have developed and validated a rapid assay for the measurement of flecainide in human plasma or serum, using a small sample volume (50 µL). METHODS: After a simple deproteination with zinc sulfate and methanol, prepared samples were injected onto a short (30 mm) analytical column and eluted using a rapid gradient elution. Detection was performed using time-of-flight mass spectrometry. Flecainide was quantified using flecainide-D4 as internal standard, with both compounds extracted from the total ion chromatogram using a ±5 ppm extraction window based on the theoretical m/z values for the protonated ions. RESULTS: The assay was linear over a putative therapeutic range (100-1500 mcg/L). Between- and within-assay imprecision and accuracy were <4.6% and 94.8%-110.0%, respectively. Matrix effects were minimal and were compensated for by flecainide-D4. There were no effects due to hemolysis or lipemia, and no carryover was apparent. Total analysis time was just 1.2 minutes (72 seconds). CONCLUSIONS: We have developed and validated a rapid method for the analysis of flecainide. The method is particularly suited for flecainide therapeutic drug monitoring, when analyzing samples from mothers receiving flecainide for the treatment of fetal tachycardia.
No takes yet. Share an insight, caveat, or question.
Rahman et al. (2018) studied Fetal tachycardia (context for therapeutic drug monitoring). Rapid assay for flecainide measurement using LC-TOF-MS was evaluated on Assay linearity, imprecision, accuracy, and analysis time. A rapid liquid chromatography and time-of-flight mass spectrometry assay for flecainide was validated, demonstrating linearity over 100-/L with a total analysis time of 1.2 minutes.
Synapse has enriched 10 related papers. Consider them for comparative context: