Here we show that a simple paper-based electrochemical sensor, fabricated by paper folding, is able to detect a 30-base nucleotide sequence characteristic of DNA from the hepatitis B virus (HBV) with a detection limit of 85 pM. This device is based on design principles we have reported previously for detecting proteins via a metalloimmunoassay. It has four desirable attributes. First, its design combines simple origami (paper folding) assembly, the open structure of a hollow-channel paper analytical device to accommodate micrometer-scale particles, and a convenient slip layer for timing incubation steps. Second, two stages of amplification are achieved: silver nanoparticle labels provide a maximum amplification factor of 250 000 and magnetic microbeads, which are mobile solid-phase supports for the capture probes, are concentrated at a detection electrode and provide an additional ∼25-fold amplification. Third, there are no enzymes or antibodies used in the assay, thereby increasing its speed, stability, and robustness. Fourth, only a single sample incubation step is required before detection is initiated.