BACKGROUND: Nocturnal hypoglycemia is a significant problem. From 50% to 75% of hypoglycemia seizures occur at night. Despite the development of real-time glucose sensors (real-time continuous glucose monitor [CGM]) with hypoglycemic alarms, many patients sleep through these alarms. The goal of this pilot study was to assess the feasibility using a real-time CGM to discontinue insulin pump therapy when hypoglycemia was predicted. METHODS: Twenty-two subjects with type 1 diabetes had two daytime admissions to a clinical research center. On the first admission their basal insulin was increased until their blood glucose level was <60 mg/dL. On the second admission hypoglycemic prediction algorithms were tested to determine if hypoglycemia was prevented by a 90-min pump shutoff and to determine if the pump shutoff resulted in rebound hyperglycemia. RESULTS: Using a statistical prediction algorithm with an 80 mg/dL threshold and a 30-min projection horizon, hypoglycemia was prevented 60% of the time. Using a linear prediction algorithm with an 80 mg/dL threshold and a 45-min prediction horizon, hypoglycemia was prevented 80% of the time. There was no rebound hyperglycemia following pump suspension. CONCLUSIONS: Further development of algorithms is needed to prevent all episodes of hypoglycemia from occurring.
No takes yet. Share an insight, caveat, or question.
Buckingham et al. (2009) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: