Electromagnetic fields from industrial and medical devices require standard safety distances for active implantable medical device users of up to 180 cm for electric fields and 450 cm for magnetic fields.
Observational (n=127)
Yes
Electromagnetic fields from industrial and medical devices can interfere with active implantable medical devices up to several meters away, requiring individual safety distance evaluations for workers.
Alerts clinicians to enforce EMF distance precautions for active implantable device patients; leaves open need for prospective clinical validation.
OBJECTIVES: Electromagnetic fields (EMF) may cause malfunctions in electronic devices, in particular in active implantable medical devices (AIMD), along with discomfort or health hazards to users. The use of AIMD by workers is increasing (especially cardiac pacemakers, implantable cardioverter defibrillators and wearable insulin infusion pumps). Electromagnetic fields may be much stronger in the working environment than applied in basic immunity tests of AIMD (based on EN 60601- 1-2:2015 and EN 50527-1:2016). European Directive 2013/35/EU regarding the safety of workers exposed to EMF considered the AIMD users to be "workers at particular risk" who need an individual evaluation of EMF hazards. The study aimed at evaluating the safety of users of AIMD in medical and industrial working environments exposed to EMF. MATERIAL AND METHODS: Near the common sources of strong EMF applied in medical and industrial use, the "standard safety distances" (SSD) for AIMD users were evaluated (i.e., distances from the EMF source, where exposure drops below limits from Recommendation 1999/519/EC and AIMD safety may be expected). The analysis is based on the results of measurements of magnetic and electric field strengths near 127 typical devices, in their normal use. RESULTS: The longest electric field related SSD was identified near dielectric sealers (up to 180 cm), and the longest magnetic field related SSD - near induction heaters (up to 450 cm). CONCLUSIONS: Electromagnetic fields related AIMD malfunctions need to be considered up to several meters from EMF sources. The "individual safety distance," that is sufficient to ensure the safety of a particular AIMD user may be significantly different (usually shorter) from the presented SSD, but needs to be considered in the context of detailed safety data from the AIMD manufacturer (if available). The labelling indicating the location of the area of a strong EMF increases safety of AIMD users in the work environment. Int J Occup Med Environ Health 2018;31(6):795-808.
No takes yet. Share an insight, caveat, or question.
Zradziński et al. (2018) conducted an observational in Active implantable medical device (AIMD) users exposed to electromagnetic fields (n=127). Electromagnetic fields (EMF) was evaluated on Standard safety distances (SSD) for AIMD users. Electromagnetic fields from industrial and medical devices require standard safety distances for active implantable medical device users of up to 180 cm for electric fields and 450 cm for magnetic fields.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: