The aim of this study was to introduce design optimization of an endosseous dental implant in order to minimize cortical bone strains at the crest of the implant/bone interface. Linear elastic p-version finite element analysis was used to find the optimal design in four different two-dimensional finite element models of the mandibular bone: model 1 (medium density cancellous bone and 2 mm thickness of cortical bone), model 2 (low density cancellous bone and 2 mm thickness of cortical bone), model 3 (medium density cancellous bone and 1 mm thickness of cortical bone), model 4 (low density cancellous bone and 1 mm thickness of cortical bone). Shape of the implant was evaluated with respect to its length and diameter, as well as the presence or absence of taper in the neck. Implants with maximum length, maximum diameter, and no taper had the lowest crestal strains in models 1, 3, and 4. Taper in the neck of the implant increased the strains in models 1, 3, and 4. In cases of poor quality of bone, increasing the diameter of the implant appeared to be the most effective means of reducing crestal bone strains.
No takes yet. Share an insight, caveat, or question.
Petrie et al. (2002) studied this question.