Answers are grounded in this paper and cite their sources.
The integration of the VMM3a Application-Specific Integrated Circuit (ASIC) into RD51’s Scalable Readout System (SRS) provides a versatile tool for the readout of Micro-Pattern Gaseous Detectors (MPGDs). With its self-triggered high-rate readout, its analogue part that allows to get information on the deposited energy in the detector, and its so-called neighbouring-logic that allows to recover information on the charge distribution, this new system has features of particular interest for digital X-ray imaging. In the present article, we want to emphasise the capabilities of VMM3a/SRS by presenting results of X-ray imaging studies. We will highlight the advantages on the energy and the spatial resolution provided by the neighbouring-logic. In the first part, we focus on spatial resolution studies. We show how segmented readout structures introduce a repeating pattern in the distribution of the reconstructed positions (using the centre-of-gravity method) and how this behaviour can be mitigated with the neighbouring-logic. As part of these studies, we explore as well an alternative position reconstruction algorithm. In the second part of the article, we present the energy resolution studies.
No takes yet. Share an insight, caveat, or question.
Scharenberg et al. (2021) studied this question.