Accidental hypothermia is a time-critical condition that requires well-versed rescuers to provide proper treatment and care in prehospital settings. However, current training procedures lack objective and quantifiable metrics for hypothermia, making it challenging to train rescuers effectively. Furthermore, relying on human actors or real patients for training can be challenging and risky in this context. An alternative is using existing simulators and manikins, yet these lack the ability to monitor heat loss and physiological responses to a decrease in body temperature. This paper addresses these limitations by introducing a novel sensor system for monitoring heat loss in medical simulators. The sensor is small, modular, and reacts to changes in the environment, providing feedback on different actions and environmental factors that would impact a real patient. Comparative testing with existing literature on human subjects demonstrates that the sensor can replicate thermal behaviour similar to real patients. Recreating existing experiments from literature performed on human subjects demonstrated that the sensor module accurately replicates the thermal behaviour of patients. The paper also outlines the iterative prototyping, testing and data-driven design process that led to the creation of this proof-of-concept.