This is the authors' abstract. We don't add key points for this paper.
Low-energy photons are necessary to improve portal image quality, but are absorbed by the linear accelerator's head components such as the target or the flattening filter. This study is a preliminary study on using low-atomic number (Z) target materials to generate photon beams with energies in the diagnostic energy range of 25 to 150 keV in the 6 MeV electron mode. Beryllium (Z = 2), carbon (Z = 4), and aluminium (Z = 13) were used as low-Z target materials in this study. GEANT4 Monte Carlo simulations were carried out to calculate the photon energy spectrum and the number of produced particles (electrons, photons, 25 to 150 keV photons) for different target materials and thicknesses. Based on the results of GEANT4 Monte Carlo simulations, photons with energies in the range from 25 to 150 keV accounted for more than 34% of the total numbers of photons in the three target materials. On the other hand, 6 MV photons, which were used to acquire the conventional portal imaging, have low-energy photons accounting for 0.3% of the total. This result means that the photons newly generated by using the low-Z target material may improve the portal image quality.