Answers are grounded in this paper and cite their sources.
A disposable amperometric biosensor for the determination of anticholinesterase activity in soil extracts is described. The sensitive membrane was obtained by co-crosslinking acetylcholinesterase and choline oxidase with bovine serum albumin using glutaraldehyde. The anticholinesterase activity of the soil extract was measured using chronoamperometry at 650 mV vs. Ag/AgCl to monitor the biocatalytically produced H2O2 before and after the inhibition step. An inhibition percentage of 38 +/- 4% was recorded for soil extracts spiked with 10 ppb of ethyl parathion. The device has the potential to be used as a gross sensor for the assessment of anticholinesterase activity in soil extracts.
No takes yet. Share an insight, caveat, or question.
Guerrieri et al. (2001) studied this question.