Answers are grounded in this paper and cite their sources.
Eleven laboratories participated in an interlaboratory study of an enzyme-linked immunoassay for limonin in grapefruit Juice. Participating laboratories received training and practice samples until familiar with the procedure. Laboratories then received 8 sample pairs of grapefruit Juice as blind duplicates. Samples were refrigerated and analyzed within 4 days of receipt. Results from 1 laboratory and 6 Individual samples were excluded from the statistical analysis. Method performance was measured by the one-way component of variance analysis. Repeatability relative standard deviations (RSDr) ranged from 7.0 to 15.6%. Reproducibility relative standard deviations (RSDR) ranged from 12.9 to 29.4%
No takes yet. Share an insight, caveat, or question.
Widmer et al. (1991) studied this question.