Key result
Higher bioelectrical impedance total body water is linked to ~33% higher odds of device fluid alerts.
Why the study?
Accurate assessment of volume status is essential in HF, and the association between CIED-derived intrathoracic impedance fluid monitoring and bioelectrical impedance analysis for systemic volume assessment required evaluation.
Does bioelectrical impedance analysis-derived total body water accurately detect and correlate with CIED-based intrathoracic impedance fluid alerts in patients with HFrEF?
Cross-Sectional (n=57)
No
Does bioelectrical impedance analysis-derived total body water accurately detect and correlate with CIED-based intrathoracic impedance fluid alerts in patients with HFrEF?
Noninvasive bioelectrical impedance analysis (BIA)-derived total body water strongly correlates with and independently predicts CIED-detected intrathoracic fluid alerts in HFrEF patients, offering a practical complementary tool for systemic volume assessment.
Answers are grounded in this paper and cite their sources.
Supports BIA for complementary fluid monitoring in HFrEF; leaves open prospective validation before clinical adoption.
BACKGROUND: Accurate assessment of volume status is essential in heart failure (HF). This study evaluated the association between cardiac implantable electronic device (CIED)-derived fluid monitoring and bioelectrical impedance analysis (BIA) for systemic volume assessment. METHODS: In this prospective cross-sectional study, 57 patients with HF carrying implantable cardioverter-defibrillators or cardiac resynchronization therapy devices with intrathoracic impedance-based fluid monitoring were enrolled. Patients were classified according to the presence of device-detected fluid alerts. Body composition and total body water (TBW) percentage were assessed by BIA (Tanita). Clinical, laboratory, echocardiographic, and device-derived data were compared. Logistic regression identified variables independently associated with fluid alerts. RESULTS: Patients with device-detected fluid alerts had higher TBW and lower body fat percentages than those without alerts (both p < 0.001). TBW showed a strong positive correlation with device-detected fluid alerts (r = 0.675, p < 0.001). A TBW threshold of 57.15% predicted fluid alerts with 80% sensitivity and 80% specificity (AUC 0.938, 95% CI 0.856-0.999; p < 0.001). Patients with fluid alerts also demonstrated more pronounced echocardiographic signs of right-sided congestion. In multivariable analysis, TBW remained the only independent predictor of device-detected fluid alerts (OR 1.33, 95% CI 1.07-1.66; p = 0.011). CONCLUSION: BIA-derived TBW was strongly associated with device-detected fluid alerts and may serve as a practical complementary method for systemic volume assessment, particularly when device-based fluid monitoring is unavailable.
No takes yet. Share an insight, caveat, or question.
Altıntepe et al. (2026) conducted a cross-sectional in Heart failure with reduced ejection fraction (HFrEF) with cardiac implantable electronic devices (CIEDs) (n=57). Bioelectrical impedance analysis (BIA)-derived total body water (TBW) percentage vs. Patients without device-detected fluid alerts (OptiVol alert negative) was evaluated on Device-detected fluid overload alert (OptiVol intrathoracic impedance threshold >60 Ohm-days) (OR 1.33, 95% CI 1.07-1.66, p=0.011). Bioelectrical impedance-derived total body water independently predicted device-detected fluid alerts with an AUC of 0.938 at a 57.15% threshold.