Why the study?
Is transvenous lead extraction effective and safe for cardiac implantable electronic device leads implanted more than 10 years prior?
Is transvenous lead extraction effective and safe for cardiac implantable electronic device leads implanted more than 10 years prior?
Transvenous lead extraction of CIED leads implanted for over 10 years is highly effective and safe, with a 95% complete extraction rate and no major complications.
Supports extraction of leads >10 years in experienced centers; extends observational data but leaves open need for prospective confirmation.
BACKGROUND: The increasing number of patients with cardiac implantable electronic devices (CIEDs) causes a rise in the absolute percentage of individuals qualifying for a transvenous lead extraction (TLE) due to infectious, vascular or lead failure related indications. As the survival time prolongs, TLE procedures more and more often concern the electrodes of long- -term functioning. Authors provide a retrospective analysis of the effectiveness and safety of TLE performed on leads implanted at least 10 years before the extraction. METHODS: Between 2008 and 2012 we performed TLE of 364 electrodes in 217 patients. Out of these, 66 (18.1%) leads in 43 (19.8%) patients had been implanted for at least 10 years. The mean dwelling time for electrodes was 161 months (120 to 330). In 62% of cases CIED-related infection was an indication for TLE. The following extracting techniques were used: manual direct traction, device traction, mechanical telescopic sheaths, autorotational cutting sheaths and femoral approach. RESULTS: Fifty-eight pacemakers and 8 defibrillating leads were extracted. Sixty-three (95%) completely, in the remaining 3 cases the clinical success was achieved with the small portion of the lead left into the vascular space. No major procedure complications were observed; minor complications were found in 3 (6%) patients. CONCLUSIONS: TLE with the use of various endovascular techniques is an effective and safe method for treating infectious, vascular and mechanical complications of long-lasting CIEDs therapy.
No takes yet. Share an insight, caveat, or question.
Maciąg et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: