The performance of a multielement array detector for measuring small spectral features superimposed on a large background has been examined. In order to have the sensitivity of detection limited only by counting statistics, two factors have to be critically considered: the nonuniform spatial response of the array detector, and the possible addition of noise by the collection system. For the special case discussed here, energy-loss spectroscopy of 100-keV electrons, the added noise was small. The detective quantum efficiency (DQE) was measured to be DQE=0.87. Several techniques reducing the effect of the nonuniform response were tested: normalizing the spectra with an experimentally measured response curve, electronic differentiation to reduce the background, dynamic gain correction, and two methods of experimentally averaging the spatial response by measuring a sequence of repositioned spectra. Preservation of the statistical information present in a representative energy-loss spectrum is shown to be feasible.