PURPOSE: The ganglion cell analysis (GCA) of the CIRRUSTM HD-OCT (Carl Zeiss, Meditec; Dublin, CA) provides measurement of the macular ganglion cell-inner plexiform layer (GCIPL) thickness. This study determined the frequency of scan artefacts and errors in GCIPL imaging in individuals undergoing HD-OCT surveillance for glaucoma. METHOD: A total of 1439 eyes from 721 subjects enrolled in a prospective study assessing predictors of glaucoma progression underwent macular GCIPL imaging with the CIRRUS HD-OCT at recruitment. The prevalence of acquisition errors, segmentation errors, and co-morbid macular pathology was determined. RESULTS: A total of 87 (6.0%) of the 1439 scans had either acquisition errors, segmentation artefacts, or other macular pathology. The most common co-morbid macular pathology was epiretinal membrane in 2.2% of eyes. CONCLUSION: The macular GCIPL scan was artefact free in 94% of eyes. However, epiretinal membrane and high myopia can cause scan artefact and should be considered when interpreting the results.